Saebo’s Care and Cleaning RecommendationsUpdated 2 months ago

The following are suggestions for cleaning Saebo products. Please refer to your infection control policy or coordinator for your required cleaning procedure.

All information below is also available in this PDF download:

Product-Care-Cleaning-Guide.pdf166.7KB


Care Cleaning Guidelines for the SaeboFlex and SaeboReach

Cleaning: We recommend using Clorox 4 in One Disinfectant and Sanitizer*. To clean properly: (1) don personal safety equipment; (2) remove liners from the plastic shell; (3) hold the can 6-8 inches away from the device; (4) spray all surfaces until visibly wet, DO NOT soak; (5) let the device dry for 5 minutes; (6) disassemble the components; (7) store the device and its components. You may also use a typical hospital/clinic rehabilitation cleaning solution**. To clean properly, spray or apply the solution to a clean cloth and wipe down all parts of the device before storing it.

Note: SaeboFlex and SaeboReach foam liners are closed-celled and are reusable after wiping. If the foam liner looks to be breaking down, usually along the edges, it is no longer closed-cell and needs to be removed (pre-fabricated shells only). Replace this liner with a new liner. Always wash and dry the client's affected hand before every use.

When to replace: Replace and discard the old components when one of the following occurs: (a) when the foam liners do not have smooth surfaces; the closed-cell design is no longer safe for the recommended cleaning (pre-fabricated shells only); (b) when the straps for the digit caps begin to fray; the straps are no longer safe for the recommended cleaning; (c) if the plastic plate or shell cracks, they are no longer safe for usage (rare). Replace these components to ensure quality infection control.

Care Cleaning Guidelines for the SaeboStretch

Cleaning: We recommend using Clorox 4 in One Disinfectant and Sanitizer*. To clean properly: (1) don personal safety equipment; (2) unzip and remove the splint cover; (3) hold the can 6-8 inches away from the device; (4) spray all fabric and metal surfaces until visibly wet, DO NOT soak; (5) let the device dry for 5 minutes; (6) reassemble and store the device.

Note: To maintain your SaeboStretch cover in good working condition, thoroughly wash and dry the client's affected hand before every use. Use soap and lukewarm water to wash the material when soiled.

When to replace: Order and replace the SaeboStretch components when the following occurs: (a) adequate amounts of pilling (small fabric balls) become present; (b) the fabric is torn; (c) the 3MTM Hook and Loop Fastener does not hold firmly; (d) the zipper does not fully zip; (e) stitching is unraveling; (f) visible discoloration is permanent, or (g) if the internal metal splint breaks.

Care Cleaning Guidelines for the SaeboGlove

Cleaning: We recommend using Clorox 4 in One Disinfectant and Sanitizer*. To clean properly: (1) don personal safety equipment; (2) separate the glove from the splint; (3) hold the can 6-8 inches away from the device; (4) spray all surfaces of the glove and splint until visibly wet, DO NOT soak; (5) let the device dry for 5 minutes; (6) remove and store all tensioners; (7) store the device.

Note: To maintain your SaeboGlove liner in good working condition, thoroughly wash and dry the client's affected hand before every use. Use soap and lukewarm water to wash the material when soiled.

When to replace: Order and replace the SaeboGlove components when the following occurs: (a) adequate amounts of pilling (small fabric balls) become present; (b) the fabric is torn; (c) the 3MTM Hook and Loop Fastener does not hold firmly; (d) stitching is unraveling; (e) visible discoloration is permanent, or (f) the wrist splint cracks or brakes.

Care Cleaning Guidelines for the SaeboStim Micro

Cleaning: We recommend using Clorox 4 in One Disinfectant and Sanitizer*. To clean properly: (1) don personal safety equipment; (2) disconnect the lead wires from the glove and the elbow sleeve; (3) separate the electrical stimulation unit from the wrist strap and place aside; (4) hold the can 6-8 inches away from the glove, sleeve, and strap; (5) spray all surfaces of the components until visibly wet, DO NOT soak; (6) let them dry for 5 minutes; (7) spray or apply the solution to a clean cloth; (8) wipe down all parts of the electrical stimulation unit and lead wires; (9) let them dry for 5 minutes; (10) store the device.

Note: Avoid spilling liquids on the SaeboStim Micro. Avoid caustic cleansers. Use soap and lukewarm water to wash the glove and elbow sleeve when soiled.

When to replace: Order and replace the SaeboStim Micro components when the following occurs: (a) adequate amounts of pilling (small fabric balls) become present; (b) the fabric is torn; (c) stitching is unraveling; (d) visible discoloration is permanent, (e) the hook and loop fasteners do not hold firmly; (f) the sheathing of the lead wires cracks or breaks; (g) the screen of the electrical stimulation unit becomes opaque; or (h) the electrical stimulation unit is not functioning properly.

Care Cleaning Guidelines for the SaeboStim One

The white controller unit, wings, and polybag should be cleaned regularly. Do not allow the interior of the controller unit to become wet during cleaning. The controller unit is fitted with a fluid detection device.

Cleaning: We recommend using Clorox 4 in One Disinfectant and Sanitizer*. To clean properly: (1) don personal safety equipment; (2) disconnect the wings and remove/cover the gel pads if needed (see Note below); (3) spray the cloth until damp, DO NOT soak cloth; (4) wipe down all parts of the controller unit, the BLUE side of the wings; and the polybag; (5) let them dry for 5 minutes; (6) turn the polybag inside out and allow to fully dry; (7) insert the wings into the polybag; (8) store the device.

Note: DO NOT attempt to take apart the wings. The gel pads of the SaeboStim One are designed to be used on one person only and not meant to be removed and stored for later application. Once the gel pads are placed onto the wings, they must stay there until ready for discard. Removal of the gel pad results in disruption of the integrity and function of the pad. Attempts to do this will result in an increased risk of client injury. Cover the gel pads with the original film and place the wings into the included polybag for later use on the same client. When removing the gel pads from the SaeboStim One, roll the pad from the proximal edge to the distal edge of the wing for easiest removal.

When to replace: Order and replace the SaeboStim One components when the following occurs: (a) the plastic shell on the electrical stimulation unit is broken; (b) the electrical stimulation unit is not functioning properly; (c) the gel pads are not sticking to the skin; (d) the wings are visibly torn or damaged; (e) the silver side of the wings is scratched.

Care Cleaning Guidelines for the SaeboStep

Cleaning: We recommend using Clorox 4 in One Disinfectant and Sanitizer*. To clean properly: (1) don personal safety equipment; (2) place the device in an upright position separating the hook and loop straps so they do not touch; (3) pull out the BOA dial; (4) pull the lace/Spectra cord completely out and lay it out in front of the device; (5) hold the can approximately 6-8 inches away from the device; (6) spray or apply the solution to all surfaces until slightly damp; DO NOT soak; (7) let the device dry for 5 minutes; (8) store the device.

When to replace: Order and replace the SaeboStep components when the following occurs: (a) adequate amounts of pilling (small fabric balls) become present; (b) the fabric is torn; (c) the 3MTM Hook and Loop Fastener does not hold firmly; (d) stitching is unraveling; (e) visible discoloration is permanent; (f) the plastic clip does not latch; (g) the BOA dial does not work; (h) the lace or spectra cord is torn; (i) the hooks or anchors are nonfunctional.

Care Cleaning Guidelines for the Saebo Mirror Box

Cleaning: We recommend using Clorox 4 in One Disinfectant and Sanitizer*. To clean properly: (1) don personal safety equipment; (2) stand up the device; (3) spray or apply the solution to all surfaces until slightly damp; (4) let them dry for 5 minutes; (5) wipe the mirror of any visual residue; (6) fold the mirror box and store the device.

Care Cleaning Guidelines for Saebo Clinical Equipment

This section includes the SaeboMAS, SaeboReJoyce, SaeboVR, and Saebo MyoTrac Infiniti.

Cleaning: We recommend using Clorox 4 in One Disinfectant and Sanitizer*. To clean properly: (1) don personal safety equipment; (2) shut off all electronics; (3) hold the can 6-8 inches away from a cloth; (4) spray the cloth until damp, DO NOT soak cloth; (5) wipe down the device on all surfaces; (6) let the device dry for 5 min; (7) safely store the device in its proper resting position.

Additional Notes

*Specific product directions can be found on the Clorox product label and website.

**Use of hospital grade cleaning products on devices besides the SaeboFlex and the SaeboReach may result in disruption of the integrity and function of the devices.